Improved modeling and quantification of chromophore concentrations in in-vivo tissue spectrophotometry

Sabine Van Huffel, Manfred Kessler and Alphons Krug (Institute of Physiology and Cardiology, University of Erlangen-Nürnberg, Germany)

Improved models based on Kubelka-Munk theory, as well as improved algorithms, for the quantification of chromophore concentrations in tissues based on optical multi-wavelength measurements, have been presented.

By means of principal component analysis (PCA) and canonical correlation analysis (CCA) on measurement spectra recorded in isolated perfused rat liver (see figure), it is shown that the currently used model describing mainly the contributions of Hb and HbO tex2html_wrap_inline1328 should be refined by including a component describing the wavelength-dependent absorption characteristics of haemoglobin-free liver. If this is the case, the inclusion of a constant term in the model function is no longer needed. The usefulness of PCA and CCA for model validation is investigated and proven by many examples. It is shown how PCA is able to detect the number of independent components with variable contribution. A subsequent CCA points out these components, which should be included in the model function.

Furthermore, three different methods for estimating the model parameters have been presented: a nonlinear least squares (NLLS) algorithm, an ordinary least squares (LS) algorithm and a mixed LS-total least squares algorithm wLS-TLS with appropriate weights. The ordinary LS algorithm is computationally most attractive but induces a too large bias in the parameter estimates which becomes dominant at low signal-to-noise ratio. In contrast, the NLLS algorithm computes the most accurate estimates due to a small bias but its computational load is 10 to 30 times (depending on the number of iterations used) higher than that of LS and wLS-TLS. The wLS-TLS algorithm offers the best compromise since it is almost as efficient as ordinary LS and almost as accurate as NLLS fitting.

It should be noted that this analysis was performed on optical measurements recorded in the visible range. The same analysis can be performed on the near-infrared measurement data and is currently being executed. Similar results are expected but, due to the lower signal-to-noise ratio, the overall accuracy of the computed chromophore concentrations will be reduced.

  figure269
Figure: 67 absorption-to-scattering spectra of isolated rat liver perfused with Hb increasing from 0.24 to 0.58 g/dl and O tex2html_wrap_inline1328 influx constant at 95%.