Metabolite-selective MRS data quantification with frequency prior knowledge

Sabine Van Huffel, Ioannis Dologlou (ESAT-MI2)

Various signal processing techniques have been proposed to improve spectral estimation of closely spaced sinusoids in the presence of noise. We focus on the problem of exponential data modeling in the presence of noise when frequency prior knowledge is available. This is a key problem in the Magnetic Resonance Spectroscopy (MRS) data fitting. Although very important, the frequency prior knowledge assumption is rather weak because it can be obtained by a simple peak picking operation on the Fast Fourier Transform (FFT) of the data under consideration. We exploit frequency prior-knowledge information to extract single peaks in MR spectra, corresponding to metabolites of interest, by means of a highly selective FIR notch filter with maximum gain at the frequency of the sinusoid under consideration. Thereafter the estimation of the parameters of the peaks is carried out using an SVD-based method known as HTLS.

The new technique improves the performance of fully automated quantification when frequency prior-knowledge is available, as shown in the figure and the table below. In addition, the proposed method is shown to be robust to prior-knowledge inaccuracies of the order of tex2html_wrap_inline1346 .

  figure208
Figure 2: Fast Fourier Transform of the simulated tex2html_wrap_inline1326 P NMR signal.

 
Method peak 1 peak 2 peak 3 peak 8
HTLS 69 96 63 60
FI-HTLS 96 100 90 65
Table 1: Number of times (out of 100 trials) that HTLS and the new method (called FI-HTLS) are able to resolve peaks 1,2,3, and 8 of the signal in the figure with added noise. The st.dev. tex2html_wrap_inline1314 for peaks 1,2 and 3 and tex2html_wrap_inline1316 for peak 8.

 
peakMethod tex2html_wrap_inline1318 (Hz) tex2html_wrap_inline1320 (Hz) tex2html_wrap_inline1322 (a.u.) tex2html_wrap_inline1324 (deg.)
1HTLS 3.40 21.13 44.45 43.46
FI-HTLS 2.53 16.86 46.90 28.27
2HTLS 2.05 31.14 129.7 29.26
FI-HTLS 1.90 21.19 83.26 25.05
3HTLS 3.69 20.60 51.63 50.68
FI-HTLS 3.00 20.82 47.71 49.00
8HTLS 0.39 4.87 23.74 5.73
FI-HTLS 0.33 4.57 23.37 4.41
Table 2: Corresponding Root mean squared error values of the parameter estimates (freq. tex2html_wrap_inline1318 , damping factors tex2html_wrap_inline1320 , amplitudes tex2html_wrap_inline1322 and phases tex2html_wrap_inline1324 ) of peaks 1, 2, 3 and 8 (excluding failures).