05/09/2025 – PhD defense – Julian Schott

System identification based detection of electrically evoked auditory steady-state responses in electroencephalography recordings for objective cochlear-implant fitting

Promotor: Marc Moonen, Jan Wouters 

Start: 05/09/2025, 1:30 pm

Location: Aula van de Tweede Hoofdwet (Room 337-01, TI Thermotechnisch Instituut, Kasteelpark Arenberg 41, 3001 Heverlee)

Abstract:

Cochlear implants (CIs) can partially restore hearing in individuals with severe to profound hearing loss by electrically stimulating the auditory nerve, thereby creating a sensation of hearing. CIs transmit essential speech information by amplitude-modulating the CI stimulation pulse trains with the envelope of incoming speech signals. This enables good speech understanding in quiet listening conditions. 

To optimize speech intelligibility and the overall hearing experience, the CI stimulation parameters need to be adjusted to the specific needs of each user. This process, known as CI fitting, is performed interactively together with a healthcare professional and relies heavily on eective communication between both parties. However, some CI users may have diculty communicating eciently, and the strong dependency of the CI fitting on the user-clinician interaction may render the fitting outcomes highly subjective. Consequently, fitting results vary between fitting sessions, and even more so between CI users and CI clinics. Reliable CI performance is therefore not guaranteed, as no standardized, evidence-based CI fitting protocol currently exists. 

One potential way to address these challenges and contribute to the development of a standardized fitting protocol is by using objective measures of hearing performance. Neural responses following CI stimulation can be recorded and used to objectively evaluate the quality of the stimulation parameters. 

In this thesis, the research conducted to enable the use of electrically evoked auditory steady-state responses (EASSRs) for objective CI fitting is presented. The EASSR, a type of neural response within the family of envelope-following responses, is elicited using amplitude-modulated CI pulse trains, and can be recorded via electroencephalography (EEG). It can provide insight into the temporal envelope encoding of CI users, and is therefore hypothesized to be valuable for objective CI fitting aimed at optimizing speech intelligibility. However, EEG recordings of CI users are contaminated with the electrical stimulation of the CI itself, making the EASSR detection challenging. This is especially the case, when the CI is operated with clinically relevant stimulation parameters (high pulse rate and monopolar stimulation mode), and when the  EEG recording channels are positioned close to the CI. 

First, a novel CI stimulation artifact removal and EASSR detection method is introduced, which is based on system identification (SI). Two independent finite impulse response (FIR) systems, of which the outputs are the EASSR and the stimulation artifact respectively, are identified using the EEG recordings. A separation of EASSR and stimulation artifact is achieved by exploiting dierences in the group delay of the identified FIR systems. The novel method was validated using EEG recordings from 16 CI users, obtained with clinically relevant stimulation parameters. The results showed very good artifact removal and EASSR detection with the SI method, even with EEG data recorded from channels closely positioned to the CI. 

Second, to further improve the clinical applicability of the SI method, a novel CI stimulation paradigm was developed that enhances artifact and response separation, and allows for group delay identification during ongoing EEG acquisition. This stimulation paradigm uses CI stimulation with increased complexity of the stimulation envelope, which facilitates artifact and response separation from a technical point of view, and allows for probing of temporal envelope encoding along dierent levels of the auditory pathway from a neurophysiological point of view. EEG data from four CI users confirms the technical advantages of the novel stimulation paradigm and validates its feasibility for use in CI users. 

Third, and finally, an objective fitting protocol using EASSRs is proposed. The protocol was tested using pre-recorded EEG data from ten CI users. Semi- objective CI maps were created and subsequently used for speech intelligibility tests. The results show good speech intelligibility with the ‘EASSR’ maps, clinically equal to the speech intelligibility as obtained with the ‘daily-life’ maps of the CI users. 

In conclusion, this thesis presents the work that enables the use of EASSRs for objective CI fitting, and demonstrates promising initial results on CI performance using objectively determined stimulation parameters. The proposed artifact removal method, along with a novel stimulation paradigm, lays the foundation for fully objective and automated CI fitting protocols based on envelope-following responses. These can be implemented in clinical routines, addressing the current challenges of CI fitting and reducing the risk of unreliable outcomes